In this study I will cover the basicss of Diagnostic Ultrasound and Elastography. The basic theory of Ultrasound such as tissue sprinkling, transducers, signal processing and show will be elaborated. Besides the jeopardies, restrictions, devices presently on the market and future applications will be covered.

Introduction

Ultrasound has been in usage as a medical imagination device for over 30 old ages, it was foremost promoted for curative applications in the 1940s to handle a assortment of conditions such as arthritis, stomachic ulcers and eczema among others. However concern was expressed over the harmful tissue effects of ultrasound which curtailed the development of diagnostic ultrasound in the undermentioned old ages. The establishing male parents of diagnostic ultrasound are Karl Theodore Dussik of Austria who published a paper on medical ultrasonics in 1942 and Professor Ian Donald of Scotland, who developed practical engineering and diagnostic applications for ultrasound in the 1950 ‘s. Finally in the late 1960 ‘s the usage of diagnostic ultrasound became widespread in infirmaries across Europe the United and States and Japan for usage in Obstetricss and Gynaecology. Unique features such as low cost, temporal declaration, not ionising radiation and portability have ensured that Ultrasound remains popular in its traditional pretense. However, current developments such as Elastography have added to the quality and applications of diagnostic ultrasound imagination. In the early 1890ss Ophir et al. , [ 1 ] introduced elastography, which is defined as biological tissue snap imagination. Primary aims of elastography were to complement B-mode ultrasound as a showing method to observe difficult countries in the chest and to look into prostate malignant neoplastic diseases, but future betterments in the engineering promises to farther heighten its value for clinical applications.

Applications

Ultrasound is used in a scope of clinical scenes including gynecology and OBs, cardiology and malignant neoplastic disease sensing. [ 2 ]

Obstetricss and Gynaecology

Detecting gestation

Measuring the size of the fetus to find the due day of the month.

Determining the place of the fetus.

Checking the place of the placenta to see if it is improperly developing over the gap to the womb.

Sing the figure of fetuss in the womb.

Checking the sex of the babe.

Checking the fetus ‘s growing rate by doing many measurings over clip.

Diagnosing tumours of the ovary and chest

Cardiology

Looking the interior of the bosom to place unnatural constructions or maps.

Measuring blood flow through the bosom and major blood vass

Urology

Measuring blood flow through the kidney.

Diagnosing kidney rocks.

Detecting prostate malignant neoplastic disease

Advantages of Ultrasound

Most ultrasound scanning is normally painless and is non-invasive.

It is widely available, easy-to-use and less expensive than other imaging methods.

Ultrasound does non utilize ionising radiation doing it particularly utile for the diagnosis and monitoring of pregnant adult females and their babes.

Ultrasound provides a clear image of soft tissues that would be ill seeable on x-ray images.

Ultrasound images are real-time, doing it an ideal tool for steering minimally invasive processs such as needle biopsies and needle aspiration. [ 3 ]

Restrictions

Supersonic moving ridges can be disrupted by air or gas ; as a consequence it is non a suited imagination technique for the intestine or variety meats obscured by the intestine. Magnetic Resonance Imaging or CT scanning would be more suited imagination techniques for these countries. As sound moving ridges pass deeper through tissue they become attenuated which makes it hard to image larger patients. Finally, Ultrasound is non effectual at perforating bone and therefore can merely see the outer surface or bony constructions, non what lies within.

Hazards to Worlds

The jeopardies associated with ultrasound are good documented. However, there have been no substantiated ill-effects of ultrasound documented in surveies in and the hazard to patients can be minimised if due safeguard is taken by the operator. The chief hazard associated with ultrasound usage is the thermic warming of tissue, which if uncontrolled can do programmed cell death. As sound energy is transmitted through tissue, some energy is reflected and some energy is lost from the supersonic moving ridge as it passes through tissue which is mostly due to viscoelastic soaking up. This loss depends on the amplitude soaking up coefficient of tissue, I± , which quantifies the loss of wave amplitude with deepness. Boness have the highest soaking up coefficient, organic structure fluids the lowest and soft tissues lie someplace in between. Most of the acoustic energy deposited is converted to heat, raising the tissue temperature. The initial rate of temperature rise is equal to: , where ( I ) is the strength of the moving ridge and ( C ) , is the specific heat capacity of the medium. [ 4 ] The strengths and powers used in modern ultrasound machines are sufficient to raise the temperature of tissues by a few grades Celsius. Heating caused by ultrasound tends to be extremely localised to the part within or instantly next to the ultrasound beam. The greatest additions in temperature occur at the surface of bone, in soft tissues conditions which could give rise to temperatures greater than 2 grades Celsius. To understate the hazard of warming, the operator should non maintain the transducer stationary for drawn-out periods and besides see the hazard of damaging tissue environing bone where temperatures will be higher.

Another hazard from ultrasound exposure is Acoustic Cavitation. This phenomenon occurs when a gas bubble in a liquid ( blood ) experiences the fluctuations in force per unit area of an acoustic moving ridge. The bubble will spread out during the reduced force per unit area and contracts during the compressive half rhythm of the moving ridge. [ 5 ] The hazard of acoustic cavitation is the mechanical forces that are exerted on the environing fluid. This could take to capillary rupture and the escape of blood contents into the environing extra-vascular infinite. [ 6 ] It is highly unlikely that acoustic cavitation occurs “ iv-vivo ” at diagnostic degrees of ultrasound. However the hazard does be for patients with coagulums or those taking anti-clotting drugs such as the acetylsalicylic acid which the operator should be cognizant of.

Finally, the presence of gas within tissue gives rise to specific mechanical effects. Any gas/tissue interface reflects all the sound energy which causes the energy denseness at the interface to duplicate and inverts the stage of the moving ridge. The deposition of energy in such construction is likely to do warming and besides an acoustic cavitation like consequence on any semi-free pits. Lung capillary harm may be caused when ultrasound scanners are used at the upper terminal of the available pulsation amplitudes. However, the hazard from gas organic structure effects is still non to the full understood and more research needs to be undertaken in this country. [ 4 ]

Principle of Operation

A transducer transmits high frequence sound pulses into the organic structure.

As the sound waves travel deeper into the organic structure they hit a boundary between tissues.

Some sound moving ridges are reflected back to the transducer while others travel farther until they are reflected back by another boundary.

Reflected moving ridges are sent back to the machines CPU.

The distance from the investigation and the tissue or organ is calculated utilizing the velocity of sound in tissue ( 1,540 ms-1 ) and the clip of each reverberations return = & gt ; d = ct/2.

This distances and strengths of the reverberations is displayed on the screen, organizing a 2D image as demonstrated in figure1 below:

Figure 1, Ultrasound image of human Foetus [ 7 ]

Echos

Sounds with a frequence above 20 KHz are supersonic since they can non be heard by the human ear. When emitted in short explosions the sound travels through media with slow contemplation coefficient and is reflected by obstructions. The sensing of this contemplation, or reverberation, of the supersonic moving ridge localises the obstruction.

Waves

As an supersonic moving ridge travels through tissue, the peak local force per unit area in the tissue increases. The oscillations of the atoms result to harmonic force per unit area fluctuations within the tissue and to a force per unit area moving ridge that propagates through the medium as neighbouring atoms move with regard to one another.

Tissue Dispersing

Tissues are composed of cells that serve as boundaries to a propagating moving ridge. As the moving ridge travels through these complex constructions, moving ridges are reflected and transmitted at any interface encountered dependent on the denseness, squeezability and soaking up of tissue at that location. The groups of cells are known as scatterers as they scatter acoustic energy. The backscattered field picked up by the transducer is used to bring forth an ultrasound image. An illustration of an ultrasound image of prostate can be seen in figure 2 below.

Fig. 2, Sonogram of prostate and it matching anatomy at the same plane [ 8 ]

The outermost bed of the prostate is shown to hold a strong reverberation, chiefly between due to the electric resistance mismatch between the environing medium ( gel ) and the prostate. The farinaceous visual aspect is called speckle [ 9 ] . It is produced by the destructive and constructive intervention of the scattered signals from constructions smaller than the wavelength ; this causes the visual aspect of bright and dark visible radiation reverberations. Therefore spot does non needfully associate to a peculiar construction in the tissue. The amplitude of spot has been represented as holding a Gaussian distribution with a certain mean and discrepancy [ 10 ] . Theses parametric quantities may be used to bespeak that the Signal to Noise Ratio of an ultrasound image is limited to 1.91 [ 10 ] . Previously, several efforts were made at speckle cancellation techniques in an attempt to increase the image quality [ 11 ] . However, spot does hold its advantages, despite being described entirely by statistics it is non a random signal, it is consistent and preserves its features when switching from place to place.

As a consequence gesture riddance techniques that can be used to find anything from blood flow to snap are possible in a technique known as spot trailing.

Figure 3, 2D Echocardiographic Right Ventricular Strain Derived from Speckle Tracking. [ 12 ]

Attenuation

As an supersonic moving ridge propagates through tissue it loses power straight relative to the distance travelled in the tissue. This phenomenon is known as fading and can be attributed to a figure of factors such as ; divergency of the wave-front, contemplation at planar interfaces, dispersing from abnormalities or point spreads and soaking up of the moving ridge energy [ 13 ] . The soaking up of moving ridge energy leads to heat addition.

Transducers

Ultrasound moving ridges are generated utilizing transducers. They typically use a piezoelectric stuff that transmits a force per unit area moving ridge when an electrical potency is applied across it. This phenomenon is reversible, that is, a piezoelectric crystal will change over an impinging force per unit area moving ridge to an electrical potency, therefore leting the same transducer to move as a receiving system. Examples of piezoelectric stuffs are polyvinylidiene fluoride, vitreous silicas and lead Zr titinate ( PZT ) . A conventional of a individual component ultrasound transducer is shown in figure 4 below.

Figure 4, typical building of a individual component transducer. [ 14 ]

The resonance frequence of the piezoelectric stuff can be given by field-grade officer = , with

regard to its thickness ( T ) and extension velocity ( degree Celsius ) .

The velocity in the PZT stuff is around 4000ms-1, so for a 5MHz transducer, the thickness should be 0.4mm midst. The fiting bed is normally coated onto the piezoelectric crystal in order to understate the electric resistance mismatch between the crystal and skin surface. To get the better of the electric resistance mismatch, the ideal electric resistance Zm and thickness diabetes mellitus of the matching bed are given by:

Zm =

and

diabetes mellitus =

with Zt = transducing electric resistance, and Z = electric resistance of the medium.

The backup beds behind the piezoelectric crystal addition the bandwidth and energy end product. If the backup bed contains air, so the air crystal intervention outputs a maximal contemplation coefficient given the high electric resistance mismatch. Besides, an air backed crystal component will hold a crystal that is comparatively un-damped, that is the signal transmitted will hold a low bandwidth and a longer continuance. The axial declaration of the transducer depends on the signal continuance or pulse breadth transmitted. Consequently, there is a trade off between transmitted power and the declaration of an ultrasound system. Therefore the type of endorsing bed to be used depends on the application. Air backed transducers are used in uninterrupted wave ultrasound applications while where high declaration images are required, to a great extent backed transducers can be used at the disbursal of decreased incursion and lower sensitiveness.

Signal Processing and Display

Figure 5, Block diagram of a pulsed-wave system and ensuing signal or image at three different stairss [ 15 ]

Figure 5 shows the different stairss used in order to get procedure and expose the signal from the tissue.

Transducer Frequency

A pulse O set continuance, frequence and bandwidth is transmitted with a trade off between incursion and declaration, therefore the frequence chosen will depend on the application. For deeper variety meats such as the bosom, liver or the uterus, frequences would be in the scope of 3-5MHz. For more superficial constructions such as the chest and thyroid or on kids, a wider scope of 4-10MHz is applied. Finally, for optic applications a scope of 7 to 15 MHz is set because of the low fading, deepness and high declaration required.

RF-Amp

The standard signal demands to be ab initio amplified so as to guarantee a good signal to resound ratio. The input of the amplifier should non hold a high electromotive force pulsation to protect the circuits while besides keeping its low noise with high addition. A typical dynamic scope at the end product would be 70-80db clip addition compensation

Time-Gain Attenuation

As the ultrasound moving ridge travels through the tissue fading occurs with increasing deepness. As a consequence, the unreal blackening of deeper constructions may happen, to avoid this, a electromotive force controlled attenuator can be used which manually adjusts the system addition consequently after an initial scan. A average fading degree with deepness is compensated for with a logarithmic electromotive force incline [ 6 ] . The dynamic scope is further reduced to 40-50db.

Compaction Amplifier

Signals are normally to be displayed on a CRT screen where the kineticss scope is typically merely 20-30db, hence as amplifier with a logarithmic response is used.

Demodulation

Since the image is in greyscale the envelope of the RF signal demands to be calculated, this is normally achieved by utilizing Hilbert transforms. The resulting signal is called an A-scan or A-mode as can be seen in figure 5.

B-Mode

The A-Scans are spatially combined after acquisition utilizing transducers and the Brightness ( B-Mode ) is created. This is the most widely used diagnostic ultrasound manner as it has a true image format. One of the greatest advantages of ultrasound imagination is existent clip scanning, which is possible due to the shallow deepness of most tissues and the high velocity of sound. Frame rates are normally in the order of 30-100MHz. The frame rate is limited to the figure of A-Mode scans acquired ( Na ) and the Maximum deepness.

Therefore:

Max Frame Rate, PRFF = [ 15 ]

M-Mode

Where tissue gesture needs to be monitored and analysed, A-Scans can be displayed as a map of clip. To accomplish this, merely one A-Scan from a peculiar tissue construction is displayed in B-Mode but followed in clip depending on the pulse repeat frequence used. This method is known as the gesture or M-Mode scan. A clip deepness show is so generated. A Typical application of M-Mode show would be in the scrutiny of the bosom valve gesture.

C-Mode

Changeless deepness manner differs from aforementioned manners due to its distinguishable usage of scanning. Alternatively of trusting on reverberations reflected from tissue, the pulsation is transmitted from one side of the organic structure by a sender and picked up on the other side by a separate transducer with a scanning gesture perpendicular to the sender beam. Applications of the C-Mode are found in superficial tissues that are comparatively homogeneous so as to guarantee travel of the reverberation through all the interfaces, for illustration in the female chest.

Diagnostic Ultrasound Machine Components

Figure 6, Basic constituents of Ultrasound machines [ 16 ]

As is demonstrated in figure 6 above, an ultrasound machine consists fundamentally of the undermentioned constituents:

Transducer investigation ; sends and receives the sound waves.

Cardinal Processing Unit ( CPU ) ; calculates and procedures signals from transducer.

Transducer pulsation controls ; changes the amplitude, frequence and continuance of the pulsations emitted from the transducer investigation.

Display ; shows image from the CPU has processed from ultrasound informations.

Keyboard ; Data input and measurement gaining control.

Disk storage device ; shops the acquired images.

Printer ; prints the image from the displayed information.

Current State of the Art – Elastography

Elasticity imagination has emerged out of ultrasound imagination in the last decennary that is based on the mechanical belongingss of tissue. Elastography can be defined as an imagination technique whereby local axial tissue strains are estimated from differential spot supplantings acquired from ultrasound frames. These supplantings are generated by a weak, quasi-static emphasis field. The end point strain image is called an elastogram.

The rule of operation of Elastography is based on two facts:

First, the mechanical belongingss of several tissue constituents can be significantly different. Figure 7 shows the cogency of this fact by demoing the scope of elastic moduli for several different normal and pathological human chest tissues. As can be seen the hardness of normal glandular tissue is different than tumerous tissue by up to one order of magnitude.

Figure 7, Elastic moduli of normal and tumorous chest tissues ; DCa: ductal carcinoma, IDCa: invasive ductal carcinoma [ 17 ]

Second, that the information contained in the spot is sufficient to picture these differences as a consequence of a quasi-static compaction. Figure 8 shows the general construct behind Elastography, tissue is insonified ( left ) before and ( right ) after a little unvarying compaction. In harder tissues ( surrounded country ) the reverberations will be less deformed than in the surrounding tissues, denoting less strain.

Figure 8, the rule of elastography. [ 18 ]

The Radio-Frequency ( RF ) supersonic informations before and after the applied compaction are acquired and speckle trailing techniques are employed in order to cipher the ensuing strain [ 1 ] . The higher the strain estimated, the softer the stuff and frailty versa. The ensuing strain image is called an elastogram. Each pel on an elastogram denotes the estimated sum of strain, Iµ , which the tissue experiences during applied compaction, defined by:

Where and denote the estimations of tissue supplanting, spaced by alteration in clip, I”t.

Elastogram Photograph Sonogram

Figure 9, eyetooth prostate. Black and white denote highest and lowest strains severally [ 1 ] .

An illustration of an elastogram is shown in figure 9. Compared with the echogram, complementary information based on the distinguishable mechanical responses and belongingss of several anatomical constructions of the prostate such as the urethra and the peripheral zones is provided. The urethral crest undergoes the highest strain given the being of the pit and higher fluid content. This demonstrates the hoe the high contrast in strain consequences in more clearly defined tissue constituents compared to a echogram.

Elastography Applications

Elastography is fast going an highly utile engineering in the diagnosing of certain malignant neoplastic diseases as strain images of soft tissues can be used to observe or sort tumors. Harmonizing to an on-going survey of the Radiological Society of North America ( RSNA ) , elastography is a technique that when added to breast ultrasound can state cancerous chest lesions from benign 1s [ 19 ] .Traditionally if a mammogram resulted in questionable findings, a patient would normally be sent for a biopsy. However, the American Cancer Society states that more than 80 per centum of chest biopsies turn out to be benign thereby seting patients under undue duress.

Prostate malignant neoplastic disease is one of the taking malignant neoplastic disease causes of work forces deceases. To better patient endurance opportunities, early sensing is critical. Early sensing of prostate malignant neoplastic disease is indispensable to supply unequivocal intervention and better patient endurance. Figure ten shows the consequences of initial clinical surveies for the prostate malignant neoplastic disease sensing.

Figure 10, leery malignant neoplastic disease parts obtained from snap imaging. [ 20 ]

Analyzing figure 10, several difficult parts of prostate are identified utilizing ultrasound snap imagination. Leery malignant neoplastic disease parts obtained utilizing snap imaging were confirmed as malignant neoplastic disease through histological analysis [ 20 ]

Devicess Presently on the Market

The major Ultrasound equipment makers are get downing to present Elastography on their premium merchandises and as upgrade options on older machines.

Siemens ACUSON S2000 Ultrasound System

Figure 11, Siemens acuson s2000 [ 21 ]

The Siemens Acuson S2000 features “ eSie Touch ” Elastography engineering.

Elastograms are formed by calculating comparative tissue distortion globally and exposing the information within a user defined part of involvement.

Axial sensing pulsations are continuously transmitted throughout the field of position to supply information about the province of tissue distortion along one axial line at a specific point in clip. Using this technique, stiff and soft tissue may be differentiated even when the tissues appear isoechoic on the B-mode test. The chief characteristics of the Siemens device are:

High declaration elastographic images may be visualised utilizing a assortment of grayscale and coloring material maps.

Shadow Measurements – Measurement callipers are automatically applied to both images in a side by side show for comparing of elastographic.

Transducer support ; eSie Touch imagination is supported on linear, endocavity and curved array transducers [ 21 ] .

Prince philips iU22 ( Vision 2010 ascent )

Figure 12, Philips iU22 [ 22 ]

As portion of an ascent on their premium diagnostic ultrasound unit, Philips Medical has introduced elastography engineering. The chief characteristics of this device are:

Highly sensitive strain based engineering which obtains elastograms from internal patient motion ( external respiration, cardiac gesture )

Distance and country measurings can be taken at the device.

Size comparings are available to formalize size and location of lesions on an elastograms.

Utilises anchoic imaging to heighten cystic constructions of the elastograms.

Colour coded strain ratio parametric image show [ 22 ] .

Future Developments

Real clip Elastography

Presently in elastography, block matching algorithms are used to gauge the supplanting and strain. Displacement calculators are based on a 2D cross-correlation method which requires a batch of generations. To expeditiously implement a realtime 2D cross correlator, processors in parallelisation are required. Field Programmable Gate Arrays ( FPGAs ) with high capacity memory Bankss, have more resources compared to general intent CPUs and digital signal processors ( DSPs ) . Therefore implementing existent clip snap imaging in FPGAs is being proposed as a feasible solution. In a paper entitled, “ FPGA based existent clip ultrasound snap imaging system ” [ 23 ] , the hardware architecture required to implement a existent clip snap imaging system utilising FPGAs is designed and tested. The method proposed is to utilize normalised cross correlativity with a 2D meat and 2D hunt for integer degree supplanting estimations. The subpixel degree supplanting is calculated by parabolic insertion in both sidelong and axial waies. The mark frame rate is at least 30 frames per second.

The Xilinx ( XC4VSX55 ) FPGA with an operating frequence of 500 Mhz is used for execution of the existent clip supplanting calculator.

Figure 13, top degree architecture for proposed FPGA based Elastography system [ 23 ]

Figure 13 shows the block diagram for a existent clip FPGA based supplanting and strain calculator. Using 2-D ( KL by KA ) meat and 2-D ( SL by SA ) hunt country, the normalised cross correlativity engine necessitating about ( KLA-KAA-SLA-SA ) A-2 generations per one end product is needed. If one frame has ( FLA-FA ) samples, there are ( FLA-FA ) A-FR end products per second where FR is the frame rate. Therefore, the sum of generations per second needed for existent clip snap imagination can be given by:

( KLA-KAA-SLA-SA ) A-2A- ( FLA-FA ) A-FR

Using 5 by 31 meats, 5 by 7 hunt scope, 256 by 4096 for frame and 30 frames per second, 340 billion MAC operations per second will be needed. However, by optimising the hardware architecture of the normalised cross correlativity engine, the impact of KA can be removed. Therefore, MAC operations were reduced to around 11 billion per second. Furthermore, to better the hardiness and quality of the supplanting estimations, two measure ( harsh and all right ) hunt and autocorrelation based mistake rectification method can be performed.

Within the given processing clip for one line, 2D cross-correlation utilizing envelope informations for harsh hunt is performed foremost, and so the radiofrequency ( RF ) information is used for both all right hunt and supplanting mistake rectification. With this system the coveted 30 frames per second was achieved [ 23 ] .

Vascular Elastography- United States Patent Application 20070282202

Changes in vas wall snap may be declarative of vas pathologies. It is known that the presence of plaque stiffens the vascular wall and plaque may take to thrombosis. It would look Elastography would be a suited engineering to name coronary artery disease, nevertheless, as tissue gesture occurs radially within the vas wall while the ultrasound beam propagates axially, gesture parametric quantities might be hard to construe. To counterbalance for this, conventional vascular elastography must be invasive ; vascular tissue is compressed by using a force from within the lms. The compaction can be induced by the normal cardiac pulsing or by utilizing a compliant intravascular angioplasty balloon. A patent has been filed for a for Non-Invasive Vascular Elastography system [ 24 ] .

The device aims to supply pre-tissue gesture and post-tissue gesture images in digital signifier of a vas delimited by a vascular wall ; the pre-tissue gesture and station tissue gesture images being representative of first and 2nd clip delayed constellation of the vas. Parts of both the pre-tissue gesture and station tissue gesture images will be partitioned into matching informations Windowss and utilizing the flight for each information window to calculate a strain tensor in each information window. The Von Mises coefficient is proposed as a new parametric quantity to characterize the vas wall. The Lagrangian spot theoretical account calculator ( LSME ) is proposed to non-invasively characterize vascular tissues because it computes the full 2D strain tensor that is required to supply the Von Mises coefficient.

Figure 14, Non-invasive vascular elastography system [ 24 ]

The block diagram in figure 14 shows the constituents of the proposed elastography device ( 10 ) . The ultrasound instrument ( 12 ) is provided with a scanhead ( 20 ) including an ultrasound transducer. The instrument ( 12 ) is coupled to an parallel to digital acquisition board ( 14 ) of a accountant ( 16 ) via a wireless frequence ( RF ) pre-amplifier ( 18 ) . The ultrasound system ( 11 ) is configured with entree to RF informations so as to let calculating vascular elastograms of vass. The ultrasound instrument ( 12 ) provides an RF end product from which the received RF informations is transferred to the pre-amplifier ( 18 ) . The acquisition board ( 14 ) allows digitising of the pre-amplified signals from the preamplifier. The accountant ( 16 ) is a Personal computer including a CPU ( 22 ) which is provided with an end product device ( 24 ) in the signifier of a show proctor coupled to the personal computing machine ( 16 ) . The accountant is provided with memory for hive awaying the scan signals and/or hive awaying elastograms information. The transducer ( 11 ) is applied on the tegument over the part of involvement and the arterial tissue is dilated by the cardiac pulsing or any other tissue dilation means.

Decision

Ultrasound has become an priceless diagnostic tool since it was widely introduced in the late 1960ss and despite the fact that it is an older imagination technique, it continues to spread out as a field and offer legion applications. In the past decennary, as faster processors have become available, new applications have emerged including contrast agents, complex transducer architecture and signal processing techniques. For the intent of this study, Elastography was chosen as the province of the art device due to the huge impact that this technique could hold on the imagination and word picture of tissue based on their mechanical properties. Evidence proves the truth of elastography at naming malignant tumerous tissue without the demand for biopsies, which averts the demand for invasive surgery and undue patient duress. For the hereafter of elastography, a assortment of really assuring methods have late developed. Crossing from manus held and existent clip application of elastography to elastic modulus maps based on the wavelength of extension through different tissues following an applied stimulation ( transeunt elastography ) and the usage of internal radiation force ensuing from the force per unit area of the beam itself to locally displace ( remote tactual exploration ) or vibrate ( ultrasound stimulated vibro-acoustography ) , to call a few.